Remote Control Commands
R&S
®
NGL200/NGM200
125
User Manual 1178.8736.02 ─ 09
OUTPut:MODE
<arg0>[, <Channel list>]
OUTPut:MODE?
[<Channel list>]
Sets or queries the output mode.
The R&S
NGL/NGM power supply series are 2 quadrant power supplies which either
source or sink current.
Parameters:
<arg0>
AUTO | SINK | SOURce | <list>
AUTO
If operates in auto mode, the R&S
NGL/NGM goes into sink or
source mode depending on the voltage across the output termi-
nal. If voltage across the output terminal exceeds the set volt-
age, current flows into the instrument, e.g. the instrument is now
operating in sink mode; vv if voltage across output terminal is
below set voltage, instrument operates as a source mode.
SINK
If operates in sink mode, current flows into the instrument. On
display, current is shown as negative current.
SOURce
If operates in source mode, current flows out from the instru-
ment.
Parameters for setting and query:
<Channel list>
<list>
Example:
OUTPut:MOD? (@1)
Queries output mode at channel 1.
Example:
OUTPut:MOD AUTO, (@1)
Set output mode to "AUTO" at channel 1.
OUTPut:SELect
<state>
OUTPut:SELect?
[<Channel list>]
Sets or queries the output state of selected channel.
Parameters:
<state>
0
Deactivates the selected channel.
1
Activates the selected channel.
*RST:
0
Parameters for setting and query:
<Channel list>
<list>
Example:
See
Example "Activating the channels"
Example:
OUTPut:SEL? (@1)
Queries output state at channel 1.
Configuration Commands