[:SOURce<hw>]:AREGenerator:RADar:BASE:ATTenuation
<AregBaseAtt>
Defines the attenuation affecting all radar objects. Together with the individual attenua-
tion for a single radar object, it forms the total attenuation for the specific object.
Parameters:
<AregBaseAtt>
integer
Base attenuation valid for all radar objects.
Range:
-50 to 150
*RST:
0
Default unit: dB
Example:
SOURce:AREGenerator:RADar:BASE:ATTenuation 50
Manual operation:
See
[:SOURce<hw>]:AREGenerator:RADar:EIRP?
Displays the measured EIRP value of the radar sensor. The EIRP value of the radar
sensor is calculated with the measured output power of the radar sensor, AREG
antenna gain, air gap. AREG internal gains and temperature compensation.
Return values:
<AregRadarEirp>
float
Example:
:SOURce:AREGenerator:RADar:EIRP
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:AREGenerator:RADar:LSENsitivity
<AregRadarLowSen>
Defines if low sensitivity is used or not.
Parameters:
<AregRadarLowSen> 0 | 1 | OFF | ON
Defines if low sensitivity is used or not. By default, low sensitivity
is used.
In this case, the saturation limit is lower, but the noise is
increased.
*RST:
1
Example:
:SOURce:AREGenerator:RADar:LSENsitivity 1
Manual operation:
See
[:SOURce<hw>]:AREGenerator:RADar:OTA:OFFSet
<AregOtaOffset>
Defines the distance (air gap) from the AREG generator to the RUT (radar under test).
SOURce Subsystem