Remote Control Commands
R&S
®
AREG100A
191
User Manual 1178.7417.02 ─ 04
Four presetting actions are available:
●
Activating the default state of all internal instrument functions (
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. GPIB address or reference oscillator source settings.
●
Activating the preset state of the parameters related to the selected signal path (
●
Activating the preset state of all parameters that are not related to the signal path
(
●
Activating the original state of delivery (factory reset,
on page 192). Only functions that are protected by a password remain unchanged
as well as the passwords themselves.
:DEVice:PRESet
Presets all parameters which are not related to the signal path, including the LF gener-
ator.
Example:
DEV:PRES
Presets all instruments settings that are not related to the signal
path.
Usage:
Event
:SOURce<hw>:PRESet
Presets all parameters which are related to the selected signal path.
Example:
SOUR:PRES
Presets all settings that are related to signal path
Usage:
Event
:SYSTem:PRESet
Triggers an instrument reset. It has the same effect as:
●
The
*RST
command
Example:
SYST:PRES
All instrument settings (also the settings that are not currently
active) are reset to their default values.
Usage:
Setting only
Preset Commands