Remote Control Commands
R&S
®
AREG100A
266
User Manual 1178.7417.02 ─ 04
:STATus:QUEue[:NEXT]?
Queries the oldest entry in the error queue and then deletes it. Positive error numbers
denote device-specific errors, and negative error numbers denote error messages
defined by SCPI. If the error queue is empty, 0 ("No error") is returned.
Return values:
<Next>
string
Example:
:STATus:QUEue?
queries the oldest entry in the error queue.
Response: 0, 'no error'
no errors have occurred since the error queue was last read out
Usage:
Query only
STATus Subsystem