Gebrauchsanweisung V04 / Operator’s Manual V04 / Manual de instrucciones V04
45
2. When Fast Mode is selected, the analyzer continues to operate in Fast Mode for the rest of the calendar day, even if the
analyzer is switched off and then on again.
3. Fast Mode is cancelled as soon as the date is automatically incremented.
!
WARNING: When performing serial measurements in Fast Mode, allow the strips to react for approxi-
mately 45 seconds before inserting them in the Urisys
1100 system and pressing START. False-low or
false-negative results may be obtained for some parameters if the reaction time is too short. Likewise,
false-high results may be obtained for some parameters if the incubation time outside the Urisys
1100
system is too long.
6.5 Strip Measurement Error
If “STRIP MEASUREMENT ERROR !” appears, the test strip and/or analyzer have probably been incorrectly used. Refer to
Section 8 for details.
6.6 Entering Patient ID, Operator ID and Authentication Password
Patient ID
When the analyzer is ready to measure (“INSERT STRIP !” displayed), you may enter a Patient ID not exceeding 13 charac-
ters in length against the currently shown sequence number by means of a barcode reader or AT/PC keyboard (see Section
9.2). The Patient ID can be verified in the display window and entered again if necessary. The last Patient ID entered is
stored when the START button is pressed, i.e. when reading commences, and is printed out and/or sent to the serial inter-
face together with the test result.
Operator ID
If the operator ID is activated, the instrument request the input of the operator ID immediately after the self check is per-
formed when the instrument is turned on, or when it leaves the stand-by mode. You may enter an operator ID up to 12
alphanumeric characters by means of a barcode reader or an AT/PC keyboard.
Authentication
In the authentication mode the user will be required to input his/her apart from the operator ID list, also the corresponding
password in order to have access to the instrument and the test results he/she had performed.
Operators with supervisor rights have access to all results, may send the instrument log file to the host or print the last
10 actions of the log file and may deactivate the authentication mode.
The list of operator IDs with corresponding passwords may be updated from the host PC using the “Download List”
function key.
!
ATTENTION: A new Patient ID should only be entered from the keyboard after the preceding measurement
has been completed and results printed out (Normal Mode). This allows the operator to check correct en-
try directly in the display.
!
ATTENTION: If a Patient ID is entered via the barcode reader while a test strip is being read, the Urisys
1100 system assigns that ID to the next sample number in the sequence. The Patient ID can only be
erased by switching the Urisys
1100 system off and then on again before starting the next reading.
!
ATTENTION: If an identification will be entered and does exceed the maximum lengths of 13 characters
for the patient Id or the 12 characters for the operator ID the cursor will move to the beginning (left side)
of the display and delete the previously written characters.