2.5 Use in custody transfer applications
This chapter contains information on how you can upgrade the ENCO08-M for
custody transfer and ensure its use here. For this purpose, seals are attached at
various points of the ENCO08-M and the housing is sealed to prevent opening.
Caution
The ENCO08-M is approved for custody transfer. It is sealed before delivery
and certain settings are locked by the approval authority.
These seals must not be damaged, destroyed or removed!
In this case, the ENCO 08 loses the calibration certification!
The ENCO 08 can only be re-certified for the in custody transfer applications
by an officially approved certification lab or an calibration official and an
additional check of the settings. The calibration officer must reattach the
seals after his verification and calibration.
The re-equipping for custody transfer is usually associated with costs