8. SPECIFICATIONS
IMSRM02-E6
93
Digital output
Failure output:
Relay contact output
Number of outputs:
1 point
Rating :
250 V AC, 0.1 A (Resistive load)
[CE/UL/CSA approved instrument: 30 V DC, 0.1 A]
Electrical life:
300,000 times or more (Rated load)
Contact type:
1a contact
Failure action:
Open at error occurrence
Digital output:
1-2
Relay contact output
Number of outputs:
4 points (PCP-B type: 2 points)
Rating :
250 V AC, 0.1 A (Resistive load)
[CE/UL/CSA approved instrument: 30 V DC, 0.1 A]
Electrical life:
300,000 times or more (Rated load)
Contact type:
1a contact
Open collector output
Number of outputs:
4 points (PCP-B type: 2 points)
Load voltage:
12 to 24 V DC
Maximum load current: 0.1 A/point, 0.8 A/common
1
Specify either relay contact output or open collector output when ordering.
2
Digital output can be selected from the following:
- Temperature alarm (alarm 1, alarm 2)
- Heater break alarm (HBA)
- Burnout alarm
- Loop break alarm (LBA)
- Temperature rise completion
Digital input (Only for PCP-B type)
Number of inputs:
3 points
Input type:
Source type
Rated input voltage:
24 V DC
Input voltage range:
21.6 to 26.4 V DC
Rated input current:
6.7 mA/point (24 V DC)
Input impedance:
3.6 k
Ω
Input operating voltage:
ON voltage:
18.5 V DC
OFF voltage:
9.0 V DC
Allocated functions:
1. Memory area transfer (8 memory areas)
2. Control RUN/STOP and memory area transfer (4 memory areas)
3. Alarm interlock release, control RUN/STOP and
memory area transfer (2 memory areas)
(Selectable)