5. COMMUNICATION DATA LIST
IMR02F04-E1
58
The communication data which will be automatically converted if the
Pressure display high/low are changed
When the data of Pressure display high (RKC communication: XV, Modbus: 00FEH) or the data of
Pressure display low (RKC communication: XW, Modbus: 00FFH) is attempted to set outside of the setting
range, these values (high or low) will be set to the Pressure display high (or Pressure display low).
In case the set value goes over the setting range: Set to the Pressure display high.
In case the set value goes below the setting range: Set to the Pressure display low.
Name Data
range
PV bias
Input span to
Input span
Alarm 1 set value to
Alarm 4 set value
Pressure display low to Pressure display high
Input error determination point (high)
Pressure display low
(5 % of input span) to
Pressure display high
(5 % of input span)
Input error determination point (low)
Pressure display low
(5 % of input span) to
Pressure display high
(5 % of input span)
Transmission output scale high
Transmission output scale low to Pressure display high
Transmission output scale low
Pressure display low to Transmission output scale high
Alarm 1 differential gap to
Alarm 4 differential gap
0 to Input span