5. COMMUNICATION DATA LIST
IMR02F04-E1
45
No. Name
RKC
Iden-
tifier
Modbus
register
address
Attri-
bute
Data range
Factory set
value
HEX
DEC
14 Digital input (DI) state
monitor
L1 00EB
235
RO
RKC
communication
Least significant digit:
The state of auto zero (DI1)
2nd digit:
The state of hold reset (DI2)
3rd digit:
The state of Interlock release
(DI3)
4rd digit to Most significant digit:
Unused
Data 0: Contact open
1: Contact closed
Modbus (Bit data)
Bit 0: The state of auto zero (DI1)
Bit 1: The state of hold reset (DI2)
Bit 2: The state of Interlock release
(DI3)
Bit 3 to Bit 15: Unused
Data
0: Contact open
1: Contact closed
[Decimal number: 0 to 7]
15 Alarm output state
monitor
Q1 00EC
236
RO
RKC
communication
Least significant digit:
The state of alarm 1 output
2nd digit:
The state of alarm 2 output
3rd digit:
The state of alarm 3 output
4th digit:
The state of alarm 4 output
5th digit to Most significant digit:
Unused
Data 0:
OFF
1:
ON
Modbus (Bit data)
Bit 0: The state of alarm 1 output
Bit 1: The state of alarm 2 output
Bit 2: The state of alarm 3 output
Bit 3: The state of alarm 4 output
Bit 4 to Bit 15: Unused
Data 0:
OFF
1:
ON
[Decimal number: 0 to 16]
16 Integrated operating time
monitor
UT
00ED
237
RO
0 to 19999 hours
Unused
00EE
238
Unused
00EF
239