2. HOST COMMUNICATION
IMR02F04-E1
14
7. Set the Device address.
Press the
key several times at input type
until device address is displayed.
Device address
InP
Add
Input type
SET
(Several times)
Press the
key to set the device address.
Setting range: 0 to 99 (RKC communication)
1 to 99 (Modbus)
As an example, factory set value “0” is set.
Add
000
0
Device address
MONI
High-lighted digit
Device address 0
8. Set the Communication speed.
Press the
key. The display goes to the
communication speed.
Communication
speed
000
0
bPS
SET
Store
Press the
key to set the communication
speed.
Setting range: 1.2: 1200 bps
2.4: 2400
bps
4.8: 4800
bps
9.6: 9600
bps
19.2: 19200 bps
38.4: 38400 bps
As an example, “38.4 (38400 bps)” is set.
bPS
9.
6
Communication
speed
MONI
38.
4
High-lighted digit
(Twice)
High-lighted digit
38.4: 38400 bps
9. Set the Data bit configuration.
Press the
key. The display goes to the data
bit configuration.
Data bit
configuration
38.
4
bIT
SET
Store
Press the
key to set the data bit configuration.
Setting range:
Refer to
Data bit configuration table
(P. 11)
.
As an example, factory set value “8n1 (Data
8-bit, Parity bit without, Stop 1-bit)” is set.
bIT
8n
1
Data bit
configuration
MONI
8n1:
Data 8-bit
Parity bit without
Stop 1-bit
High-lighted digit