61
Measurement
On this unit, PPV, VDV, and other values are calculated at one time. The
calculation results are saved in an SD card. Via the Store menu, you can select
the following three Store modes.
Manual mode
This mode allows calculation during a preset “Measurement Time.” Instan-
taneous values are not saved.
In Manual store mode, the following measurement values are saved.
PPV, D.F., KBFmax (SBR: veff, max), Acc., MTVV, VDV, C.F., Disp., PVS,
OVER / UNDER
In SBR mode, MTVV, VDV, C.F. measurement values are not saved.
Measurement
start
measurement time
Measurement
stop
Auto mode
Calculation is performed at calculation intervals during a preset “Total Mea-
surement Time.” At the same time, instantaneous values are saved every second.
In Manual store mode, the following measurement values are saved.
PPV, D.F., KBFT (SBR: veff, max, 30), Acc., MTVV, VDV, C.F., Disp., PVS,
OVER / UNDER
In SBR mode, MTVV, VDV, C.F. measurement values are not saved.
Measurement time
(actual measurement time)
set as calculation interval
Measurement
start
Data saved
in address 1
Data saved
in address 2
Measurement
in progress
Data saved
in address 3
Measurement
in progress
Total measurement time
Measurement
in progress
..........
Measurement
stop