141
Serial Interface
DRD
Continuous output
Data are sent periodically to the computer every 100 msec.
To stop the data transfer, send the stop request transfer code <SUB> (hexa-
decimal notation: 1AH).
Request command DRD
Response data
d1,d2,...,d67
Returned value
d1 = “xxx.x”
Acc. of X channel (Inst value)
d2 = “xxx.x”
PPV of X channel (Inst value)
d3 = “xxx.x”
D.F. of X channel (Inst value)
d4 = “xxx.x”
Disp. of X channel (Inst value)
d5= 0 or 1
Inst Overload information
(1: Yes, 0: No)
d6= 0 or 1
Inst Under-range information
(1: Yes, 0: No)
d7 = “xxx.x”
Acc. of X channel (Max hold
value)
d8 = “xxx.x”
PPV of X channel (Max hold
value)
d8 = “xxx.x”
D.F. of X channel (Max hold
value)
d10 = “xxx.x”
Disp. of X channel (Max hold
value)
d11 = 0 or 1
Max hold Overload information
(1: Yes, 0: No)
d12 = 0 or 1
Max hold Under-range informa-
tion
(1: Yes, 0: No)
d13 = “xxx.x”
PPV of X channel (Calc value)
d14 = “xxx.x”
D.F. of X channel (Calc value)
d15 = “xxx.x”
Disp. of X channel (Calc value)
d16 = “xxx.x”
Acc. of X channel (Calc value)
d17 = “xxx.x”
C.F. of X channel (Calc value)
d18 = “xxx.x”
MTVVof X channel (Calc value)