80
Calibration
6. Set a sample for cleaning (in general purifi ed water) then either click the “Flush”
button, or the “FLUSH” button on the front of the unit.
Important
When operating by pressing the “FLUSH” and “START”
buttons on the unit, ensure that the “KL-04 - Calibration”
window is set in an active window, in which the required
Calibration Parameter Name is displayed. If it is not in an
active window, operations may be different from the speci-
fi ed Parameters.
7. Set up pure water for the Blank (Blank is a sample of pure water without any
PSL-particles) measurement, then click the “Blank” button (displayed on the left
hand side of the same row) to select the particle size (e.g. 10.350 µm) from the
Calibration particle size list.
Upon completion of the Blank Measurement, the result will be displayed on the
right hand side of the “Blank” button.
8. Calibrate. Set sample(s) for calibration then select the specifi c required calibration
particle size (e.g. 10.350 µm) from the list.
Upon completion of the calibration task, the calibrated particle size will be dis-
played at the right of the particle size.
Note
The calibration result will be displayed in black if it is accepted,
and in red if rejected.