![Rion KL-04 Instruction Manual Download Page 19](http://html1.mh-extra.com/html/rion/kl-04/kl-04_instruction-manual_1463442019.webp)
3
Outline
In other words, if the Performance-Test data KL-04_Basic is selected, the Calibration data
KL-04_Basic is applied to the unit and the unit is in the calibrated condition based on the
method decided by RION Co., Ltd.
Important
The Performance-Test data complying with the purpose of the
quality control test must be selected. An accurate measure-
ment cannot be performed when wrong Performance-Test
data are selected.
Multiple Performance-Test parameters (condition of test, condition of acceptance or rejec-
tion) can be registered to the same Calibration data. This is useful when carrying out the
Performance-Test as an inspection before starting any measurement.
A formal Performance-Test must be implemented to judge if the condition of the unit is
appropriate. However, it is possible to confi rm the condition of the unit by setting up a
Performance-Test parameter with a name dif-
ferent from the formal Performance-Test, such
as the chart on the right, and implement the
Performance-Test when needed, such as at an
inspection before commencement. In other
words, multiple Performance-Test parameters,
such as formal Performance-Test and simple
Performance-Test, can be set to the same Cali-
bration data and used appropriately.
Important
Performance-Test data obtained with appropriate or equivalent
(to a chosen pharmacopeia) Performance-Test parameters
must be selected for a measurement. The measurement
cannot be carried out properly if the Performance-Test
data is obtained with inappropriate Performance-Test pa-
rameters.
Performance-Test
Measurement parameter
Performance-Test 1
Calibration
Simple Performance-Test
Summary of Contents for KL-04
Page 2: ......
Page 6: ...iv...
Page 10: ...viii...
Page 12: ...x...
Page 16: ...xiv...
Page 166: ...No 31947 06 11 Printed in Japan...