71
Performance-Test
Number of Pre-Measurements (times):
This enables setting of the number of repetitions of the Pre-Measure-
ment prior to the Measurement.
As the measuring sample may not necessarily be evenly distributed, It
is therefore necessary to set the inside channel to the sample number
concentration level. The Pre-Measurement is, therefore, required for
this purpose and the suction volume etc., are carried out in the same
manner as for the actual Measurement. However, the measurement
result will not be included in the calculation of the average.
The allowable setting range is between 0 to 10 times.
Number of Measurements (times):
This refers to setting the Number of Measurements. The allowable set-
ting range is between 1 to 100 times.
Drain After Every Measurement?:
Drain a sample from the syringe after every measurement.
Flush
Aspiration Flow Rate (mL/min):
This sets the suction velocity of the syringe during cleaning.
The allowable setting range is 5 to 100 mL / min (1 mL step).
Drain Flow Rate (mL/min):
Set the syringe’s Drain velocity while cleaning.
The allowable setting range is 1 to 100 mL / min (1 mL step).
Flush Volume (mL):
Set the required volume for the syringe to draw up each time during
cleaning.
The allowable setting range is between 0.2 to 25.0 mL (0.1 mL step).
Repeat (times):
Set the number of times the unit carries out cleaning. The allowable
range is between 1 to 100 times.