To display a signal automatically
The oscilloscope has an Auto feature that automatically sets up the oscilloscope to
best display the input signal. Using Auto requires signals with a frequency greater
than or equal to 50 Hz and a duty cycle greater than 1%.
When you press the AUTO button, the oscilloscope turns on and scales all channels
that have signals applied, and it selects a time base range based on the trigger
source. The trigger source selected is the lowest numbered input that has a signal
applied.
The DS5000 series are two-channel oscilloscopes with an external trigger input. In
this exercise you connect a signal to the channel 1 input.
1.
Connect a signal to the oscilloscope.
2.
Press AUTO.
When you press the AUTO button, the oscilloscope changes the front-panel setup to
display the signal. It automatically adjusts the vertical and horizontal scaling, as well
as the trigger coupling, type, position, slope, level, and mode settings.
Copyright 2004 Rigol Electronic
1-9
User Manual for DS5000 Series