RESA ring installation
Recommended taper roundness
DO
= Nominal external diameter
Recommended taper diameter (
DT
)
*
Allow 2 mm for 417 mm, 489 mm and 550 mm rings only.
5 mm Max.
DT
15° ±0.2°
7 min
1*
Recommended surface finish ≤ Ra 1.2
NOTE:
It is recommended that the mounting surface is a turned, rather than ground finish.
Taper mount
Interference fit
‘A’ section
'B' section
Not applicable
Notes
Enables simplest adjustment
Offers highest accuracy
Enables eccentricity to be compensated.
Offers excellent mechanical stability against
thermal cycling, shock and vibration.
Minimises cost of substrate preparation.
Alternative installation
Will not correct eccentricity of the
supporting shaft
Diameter
(mm)
Roundness value
(mm TIR)
< 52
0.025
150 to 255
0.050
≥ 300
0.075
DO
(mm)
DT
(mm)
52
33.85
33.65
57
40.85
40.65
75
58.85
58.65
100
83.85
83.65
103
83.85
83.65
104
83.85
83.65
115
98.85
98.65
150
133.85
133.65
200
183.85
183.65
206
189.85
189.65
DO
(mm)
DT
(mm)
209
189.85
189.65
229
212.85
212.65
255
238.85
238.65
300
283.85
283.65
350
333.85
333.65
413
395.85
395.65
417
383.85
383.65
489
454.85
454.65
550
513.85
513.65
Taper mount method
Step 1
Mounting shaft specifications
Mounting options