8
Introduction
The APCA-45 is designed to be mounted to a
suitable machine bulkhead, surface bulkhead or
panel and requires through holes to be provided to
fix it in place. Bulkhead or panel must be stiff and
not transfer excessive vibration to the APCA-45
during operation.
Preparing the machine for APCA-45 installation
Ø 9.25 (0.36) through × 3
Machine cutout
(orientation to suit installation)
Dimensions in mm (in)
10
(0.39)
37.5 (1.48)
18.75 (0.738)
16.5 (0.65)
Ø 16 (0.63) through
M4 holes × 4
10 (0.39) deep minimum
18.75
(0.738)
37.5 (1.48)
10
(0.39)
18 (0.71)
NOTE:
If the 4 M4 screw
holes are through holes, it is
recommended that a sealant
is used on these holes.
CAUTION:
The APCA-45 is designed to be
mounted horizontally or at any angle between
horizontal and vertical with the nose cone facing
down. It must NOT be mounted with the nose
cone at any angle above horizontal (see diagram
on page 9).
Summary of Contents for APCA-45
Page 1: ...Installation and user s guide H 6596 8500 01 B APCA 45 tool setting probe...
Page 2: ......
Page 4: ...This page is intentionally left blank...
Page 33: ...15 Pneumatic options continued Bulkhead Dual Air air bleed circuit 1 V3 V2 V1...
Page 35: ...17 Pneumatic options continued Dual Air air bleed circuit 2 Bulkhead V3 V2 V1...
Page 47: ......