18
Operation
sequence
Solenoid
valve 1
(V1)
Solenoid
valve 2
(V2)
Solenoid
valve 3
(V3)
End of sequence
trigger
Monitor Variable to
set
Comment
1 Probe cover
extend
1
0
0
Pressure switch
activation
Pressure
switch
2 Adjustable hold
(extended) time
0
0
0
Time
Pressure
switch
Sequence
hold time
Pressure
switch must
read >4.5 bar
(65.3 psi)
3 Air bleed
(extended) on
0
0
1
Time
Pressure
switch
Bleed time Pressure
switch must
read >4.5 bar
(65.3 psi)
4 Air bleed
(extended) off
0
0
0
Time
Pressure
switch
Sequence
hold time
Pressure
switch must
read >4.5 bar
(65.3 psi)
5 Probe cover
retract
0
1
0
APCA-45 sensor
APCA-45
sensor
6 Air bleed
(retracted) on
0
0
1
Time
APCA-45
sensor
Bleed time
7 Air bleed
(retracted) off
0
0
0
Time
APCA-45
sensor
Sequence
hold time
8 Adjustable hold
(retract) time
0
0
0
Time
APCA-45
sensor
Sequence
hold time
Operation sequence
(circuit 2)
Pneumatic options
(continued)