Table 201 – Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The
ClearChart 4
is intended for use in the electromagnetic environment specified below. The
customer or user of the
ClearChart 4
should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment
- Guidance -
RF Emissions
CISPR 11
Group 1
Class A
The ClearChart 4 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
Harmonics
IEC 61000-3-2
Class A
The ClearChart 4 is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low-voltage power sup-
ply network that supplies building for domestic power.
Flicker
IEC 61000-3-3
Complies
Note:
The emissions characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR
11 class B is normally required) this equipment might not offer adequate protection to radio-
frequency communication services. The user might need to take mitigation measures, such
as relocacting or re-orienting the equipment.
Guidance & Manufacturer’s
Declarations
13795-101 Rev. F
40