13795-101 Rev. F
38
Specifications
Catalog Numbers:
13795 (ClearChart 4), 13775 (ClearChart 4X), 13790 (AcuityChek)
Physical Dimensions
Size:
Weight, unpacked: 13.1 lbs. (5.9 Kg)
Height: 15.60 in. (39.62 cm)
Width: 23.20 in. (58.92 cm)
Depth: 3.00 in. (7.62 cm)
Electrical
Voltage: 100 - 240 volts AC 50/60 Hz
Input Power: 50-75 VA
Fuses: Time-Lag (2.5A, 250V), 5 x 20mm, RoHS
Operational Conditions
Environmental:
The environmental conditions are as follows:
Operating:
Temperature 10° C (50° F) to 35° C (95° F)
Relative Humidity: 30% to 75%
Atmospheric Pressure: 80 (23.6 in. Hg) to
106 kPa (31.3 in. Hg)
Transportation & Storage:
Temperature -20° C (-4° F) to +50° C (122° F).
Relative Humidity: 10% to 80% (non-condensing)
Atmospheric Pressure: 70 (20.7 in. Hg) to
106 kPa (31.3 in. Hg)
Exposure to extreme temperature conditions indicated above must not
exceed 15 weeks.
Expected Service Life
The expected service life for this instrument is 7 years.
Disposal
This product does not generate any environmentally hazardous residues. At the end of its product
service life, follow your local laws and ordinances regarding the proper disposal of this equipment.
Software Revision
The software revision is shown on the splash screen at start up. Users can also contact Reichert for
the software revision. The serial number identifies the manufacture date and will be used to identify
the software version.
-20°C
50°C
10%
80%
106 kPa
70 kPa