Doc-1301-Ops Rev
2.9.1
10/30/08
Refraction Technology, Inc.
79
4.9.4
Updating firmware using ftp
1.
ftp to the 130-01 using its IP address:
username: reftek
password: reftek
2.
Use the ls ftp command to list the contents of the Compact Flash™.
3.
Enter bi and ha to select Binary and Hash options.
4.
Use the put ftp command to send the file (main.s3) to the Compact Flash™
installed on the 130-01.
5.
With the Compact Flash™ with the main.s3 image installed in the 130-01 DAS,
issue a reset command from the 130-01
-OR-
Use the RTCC command and control interface program
Use the RT130-SM GUI Reset System command
6.
On power-up, the 130-01 checks the Compact Flash™ for the presence of
‘main.s3’ in the root directory.
7.
If the ‘main.s3’ file is present on the Compact Flash™, the 130-01:
(a) Reads the file.
(b) Re-programs the internal flash memory.
(c) DELETES the file.
DO NOT DISTURB THE UNIT DURING THIS PROCESS.
4.9.5
Updating firmware at the 130-01
1.
Issue a Reset command from a PDA running PFC_130 or Physically disconnect
and reconnect power to the unit.
2.
Observe the LCD for the following messages:
READING DISK DO NOT DISTURB
WRITING FLASH DO NOT DISTURB
3.
The 130-01 DAS resets and returns to normal messaging.
Summary of Contents for 130-01
Page 16: ...130 01 Operations Refraction Technology Inc 2 ...
Page 54: ......
Page 82: ...130 01 Operations Refraction Technology Inc 68 Figure 51 PTO Connector ...
Page 87: ...Doc 1301 Ops Rev 2 9 1 10 30 08 Refraction Technology Inc 73 ...
Page 88: ...130 01 Operations Refraction Technology Inc 74 ...
Page 89: ...Doc 1301 Ops Rev 2 9 1 10 30 08 Refraction Technology Inc 75 ...
Page 96: ......