11. EMC Manufacturer Declarations
Electromagnetic Compatibility (EMC) Statement for Home Healthcare Environment
The RecoveryPump
®
system has been evaluated to international standard
IEC 60601-1-2 "General requirements for basic safety and essential
performance - Collateral standard: Electromagnetic compatibility -
Requirements and tests".
The Medical Electrical Equipment needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC information
provided here within the accompanying documents.
Portable and mobile RF communication can affect the Medical Electrical
Equipment. See below recommended separation distances between portable
and mobile RF communication equipment and the RecoveryPump
®
system.
Wireless communications equipment such as wireless home network devices,
mobile phones, cordless telephones and their base stations,
walkie-talkies can affect this equipment and should be kept at least a distance
d = 3.3 m away from the system.
Rated Maximum Output
Power of Transmitter (W)
Separation Distance* according to
Frequency of Transmitter (m)
0.01
0.23
0.1
0.73
1
2.3
10
7.3
100
23
*Note
: The distance calculated from 800 MHz to 2.5 GHz
Caution:
•
Do not use a mobile phone or other devices that emit electromagnetic
fields, near the unit. This may result in incorrect operation of the unit.
•
This system should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, this
machine should be observed to verify normal operation in the
configuration in which it will be used.
L98005H1R-A
31