Model 737R - electromagnetic emissions - manufacturer declaration
The
Model 737R
is intended for use in the electromagnetic environment specified below. The customer or
the user of the
Model 737R
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The
Model 737R
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The
Model 737R
is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low
voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Model 737R - electromagnetic immunity - manufacturer declaration
The
Model 737R
is intended for use in the electromagnetic environment specified below. The customer or
the user of the
Model 737R
should assure that it is used in such an environment.
Immunity Test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Complies
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
Complies
Mains power quality should be that of a typical
domestic or hospital environment.
Surge
IEC 61000-4-5
±1 kV line to
line
Complies
Mains power quality should be that of a typical
domestic or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
>95% dip in
U
T
for 0.5 cycle
60% dip in
U
T
for 5 cycles
30% dip in
U
T
for 25 cycles
>95% dip in
U
T
for 5 cycles
Complies
Mains power quality should be that of a typical
domestic or hospital environment.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
Complies
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical domestic
l or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
32