Cleaning the RecoveryPump
®
Console
Caution:
Disconnect the console from electrical outlet before cleaning. Let it dry
completely before reconnecting it to the electrical outlet.
•
Wipe clean using a
slightly damp cloth
.
•
Do not allow any moisture or liquids to enter the console!
Cleaning the Garments
a. Never submerge the garments in liquid. They are cleanable by surface
wiping only.
b. Gently wipe the external and internal surfaces of the compression
garment using a soft cloth, warm water not exceeding 100°F (40°C) and a
mild detergent. Do not submerge in liquid! Do not allow liquid to enter the
air inlets of the garment! Wipe with clear water to remove any detergent
residue, again taking care that water does not enter the air inlets of the
garment.
c. Wipe dry using a soft towel only.
d. When dry, the external and internal surfaces of the compression garment
may be wiped down with alcohol. When using alcohol, wear gloves and
work in a well-ventilated area.
e. Allow the garment to air dry completely before use.
Caution:
•
Do not hand or machine wash. Surface wipe only!
•
Do not allow liquid to get into the air inlets.
•
Do not use bleach.
•
Do not dry clean.
•
Do not wring, iron, tumble or force heat dry.
Storage and Transportation
a. Keep the console, hoses and garments away from direct sunlight.
b. Store in a dry, shaded place (see caution below).
c. Do not twist or fold the hose bundle.
d. Store the power cord lightly wrapped and secured.
Caution:
The console can be transported or stored for short periods of time within:
•
Temperature range of -4°F to 158°F (-20°C to 70°C)
•
Humidity range of 10%rH to 93%rH non-condensing
•
Atmospheric pressure range of 190hPa to 1060hPa.
Allow the console to reach reasonable room temperature 50°F to 86°F
(10°C to 30°C) before operating.
28
L98005H1R-A