MT5 Multi-Tracer Operating Manual
page 26
•
Off
– Keep this module’s load turned off. Therefore the module will operate in an
open-circuit condition.
•
Peak Power
– Operate the module in a peak-power tracking mode (see Appendix B
“Peak-Power” tracking for details).
•
Fixed Voltage
– This mode operates the module at the fixed voltage given. Enter the
voltage in the associated box.
Enable Auto IV Curves
This check box enables this module for automatic IV curves. See the section 4.3.5
“IV
Curves Setup” for additional details. This setting determines if an IV curve is taken
automatically for this module
when the “Global IV Curve Mode” is set to “Channel
Defined”. See the section 4.3.1 “Global Load & IV Curve Control” for details on
overriding this setting.
Integration Period
The “Integration Period” setting defines the number of microseconds used for module
voltage and current data sampling. The default 100 uS is recommended and should not
be changed unless necessary. Larger values will help reduce signal noise by averaging
the signal over a longer period.