MT5 Multi-Tracer Operating Manual
page 87
Current Measurements
Measurements
Range
Resolution
*Accuracy
Temp Co.
20 A Full Scale
20.0 A
610 uA
0.1% + 0.02%
50 ppm/°C
2.0 A
61 uA
0.2% + 0.04%
50 ppm/°C
15 A Full Scale
15.0 A
460 uA
0.1% + 0.02%
50 ppm/°C
1.5 A
46 uA
0.2% + 0.04%
50 ppm/°C
12 A Full Scale
12.0 A
366 uA
0.1% + 0.02%
50 ppm/°C
1.2 A
37 uA
0.2% + 0.04%
50 ppm/°C
10 A Full Scale
10.0 A
300 uA
0.1% + 0.02%
50 ppm/°C
1.0 A
30 uA
0.2% + 0.04%
50 ppm/°C
8 A Full Scale
8.0 A
245 uA
0.1% + 0.02%
50 ppm/°C
0.8 A
25 uA
0.2% + 0.04%
50 ppm/°C
5 A Full Scale
5.0 A
150 uA
0.1% + 0.02%
50 ppm/°C
0.5 A
15 uA
0.2% + 0.04%
50 ppm/°C
2 A Full Scale
2.0 A
60 uA
0.1% + 0.02%
50 ppm/°C
0.2 A
6 uA
0.2% + 0.04%
50 ppm/°C
1 A Full Scale
1.0 A
30 uA
0.1% + 0.02%
50 ppm/°C
0.1 A
3 uA
0.2% + 0.04%
50 ppm/°C
* Percent of R Percent of Range @ 25 °C
Auxiliary Inputs
Common
Measurements
Range
Mode
Resolution
*Accuracy
Temp Co.
General Purpose DC
±5V
±5V
150 uV
0.1% + 0.02%
50 ppm/°C
±500
mV
±5V
15 uV
0.1% + 0.02%
50 ppm/°C
±50 mV
±5V
1.5 uV
0.2% + 0.04%
50 ppm/°C
±5 mV
±5V
0.15 uV
0.3% + 0.06%
50 ppm/°C
* Percent of R Percent of Range @ 25 °C
Temperature Inputs
Measurements
Range
*Resolution
Accuracy
Temp Co.
Type T Thermocouple
-40 to 200 C
0.06 C
+/- 1.5 C
50 ppm/°C
Type K Thermocouple
-40 to 200 C
0.08 C
+/- 2.0 C
50 ppm/°C
* Worst case resolution
Specifications subject to change without notice.