Trying out the instrument
R&S
®
SMCV100B
45
Getting Started 1432.7046.02 ─ 04
● The "User 2" connector is an input for the "Global Trigger" signal.
The LED next to the connector is green.
Find the physical location of each connector
Use the built-in "Trigger Marker Clock > Show " function to display the loca-
tion of the selected connector. A blinking marker on the front/rear panel view
also indicates the selected connector.
To reconfigure the trigger settings
We assume that the instrument is configured as described in
erating a digitally modulated signal"
on page 40 and the default connector map-
).
1. In the block diagram, select "Baseband > Misc > Custom Digital Mod > Trigger
In".
2. Select the following settings:
a) "Mode > Armed Auto"
b) "Source > External Global Trigger".
3. Select "Global Connector Settings > Routing".
4. For "User 2", select "Direction > Input" and "Signal > Global Trigger".
Triggering the instrument with an external signal