Operating Basics
R&S
®
NGL200/NGM200
29
User Manual 1178.8736.02 ─ 06
Function
Description
"Output Delay"
If enabled, the icon is highlighted in white.
The delay is the time between activation of the out-
put and applying voltage to the output.
Sense connection
If sense connection is detected, the icon is highligh-
ted in white.
5.1.2
Channel Display Area
The R&S
NGL/NGM displays two channels display area (Ch1, Ch2) for NGL202,
NGM202 and a single channel display area (Ch1) for NGL201, NGM201. The respec-
tive channel settings and functions are displayed for each channel.
Figure 5-2: Channel display area for 2-channel model
1 = Output power displays in watt
2 = Source output resistance/emulated internal impedance displays in ohms
3 = "Settings" button opens instrument device/channel menu window
4 = "Expand/Collapse" button toggles between home window and channel overview window
5 = Output voltage displays in volt. The display resolution for voltage is five digits after the decimal point
6 = Set voltage level. This level is limited by the set value defined in "Safety Limits"
7 = Output current displays in ampere. The display resolution for current is five digits after the decimal point
8 = Set current level. This level is limited by the set value defined in "Safety Limits"
9 = DVM function (available only with R&S NGM power supply series)
Operating mode
Different font colors on the screen are used to differentiate the various output status
and operating condition of the instrument. It is easy to know and confirm the different
output status and operating conditions of the instrument by looking at the colors.
Display Overview