Remote Control Commands
R&S
®
NGL200/NGM200
137
User Manual 1178.8736.02 ─ 06
Example:
DIO:FAUL:CHAN 1
Sets channel 1 as the output fault source.
DIO:FAULt:SOURce
<arg0>
Sets or queries the "operation modes" of the digital output fault source
.
Parameters:
<arg0>
CC | CV | CR | SINK | PROTection
Example:
DIO:FAUL:SOUR PROT
Sets the "operation modes" of the digital output fault source as
protection. If any of the protection modes (OCP, OCP, PPP, OTP
and Sense) of the digital output fault source is active, a trigger
event is triggered to the trigger-out signals on the instrument.
DIO:FAULt[:STATe]
<arg0>
Enables or disables the digital output fault.
Parameters:
<arg0>
1 | 0
Example:
DIO:FAUL 1
Enables the digital output fault.
DIO:OUTPut:SOURce
<arg0>, <arg1>
DIO:OUTPut:SOURce?
<arg0>
Enables or disables the digital output source.
.
Parameters:
<arg1>
OUTPut | TRIGger | FORCed
OUTPut
Selected channel output is used as the digital output source.
TRIGger
Selected channel external trigger signal is used as the digital
output source.
FORCed
Selected output is forced to high level and can be switched by
on page 138 command.
Parameters for setting and query:
<arg0>
1 | 2
Channel selection for the digital output source.
Advanced Operating Commands