Remote Control Commands
R&S
®
NGL200/NGM200
116
User Manual 1178.8736.02 ─ 06
7.5.8
Range/DVM Setting
The DVM and range settings are available only with R&S NGM power supply series
equipped with R&S NM-K104 (P/N: 3643.9927.02).
The
SOUR:VOLT:DVM
contains commands for activating the DVM function.
The
SENSE:CURR:RANG
and
SENSE:VOLT:RANG
contains commands for setting the
voltage range and current range of the measurements.
All these commands require
*OPC?
at the end of the command execution.
....................................................................................116
..................................................................................... 116
...................................................................................117
......................................................................................117
................................................................................... 117
[SOURce:]VOLTage:DVM[:STATe]
<arg0>
Enables or disables digital voltmeter measurements.
Parameters:
<arg0>
1 | 0
1
Enables digital voltmeter measurement.
0
Disables digital voltmeter measurement.
Example:
VOLT:DVM 1
MEAS:VOLT?
-> 1.000E+00
Enables and returns digital voltmeter measurement.
SENSe:CURRent:RANGe:AUTO
<arg0>
Enables or disables auto range for current measurement accuracy.
Parameters:
<arg0>
1
Enables auto range for current.
0
Disables auto range for current.
Example:
SENS:CURR:RANG:AUTO 1
Enables auto range for current.
Configuration Commands