Instrument Functions
R&S
®
NGL200/NGM200
43
User Manual 1178.8736.02 ─ 06
Figure 6-4: Constant resistance dialog
6.2.2
Fast Transient Response
With fast transient response, the R&S
NGL/NGM is able to quickly stabilize the output
voltage upon a step change in the load current.
Load transient recovery time can be switched between 30
μ
s ("Fast Transient
Response" = "ON") or 100
μ
s ("Fast Transient Response" = "OFF") for the output volt-
age to recover and stay within 20 mV of the nominal output voltage follows by a step
change in the load current.
Figure 6-5: Transient response graph
1. Press [Settings] key.
The R&S
NGL/NGM displays the device/channel menu window.
2. Select the required channel tab to apply fast transient response.
The R&S
NGL/NGM displays the selected channel menu.
Activating the Channels Output