Instrument Functions
R&S
®
NGL200/NGM200
70
User Manual 1178.8736.02 ─ 06
The R&S
NGL/NGM activates the logging and disables the settings for file saved
location and logging mode settings.
9. Applicable only with R&S NGL-K103, activate the "Triggered" menu item if data
logging is required under triggered conditions.
If activated, the R&S
NGL/NGM executes the data logging if a trigger event occurs.
and
.
10. Configure the "CSV Settings".
6.11
FastLog
Available only with the NGM models.
A USB stick with a minimum of 10MB/s continuous write speed is recommended for
data storage.
The FastLog can write the measurement data (votage/current) either to the USB
device or to a SCPI client. In SCPI mode the client is required to be connected to the
R&S NGM before "FastLog" is enabled
"FastLog" is a much faster data logging for voltage and current measurements. This
fast logging features the benefit to capture short peak changes in voltage and/or cur-
rent measurements which is particular useful in the measuring of low energy device,
e.g. IoT products.
Refer to the data specification for the maximum data logging and resolution for NGM
models.
1. Press [Settings] key.
The R&S NGM displays the device/channel menu window.
2. Select the required channel tab to configure "FastLog" function.
The R&S NGM displays the selected channel menu.
3. Select "FastLog" menu item from the menu.
The R&S NGM displays the "FastLog" dialog.
FastLog