Operating Basics
R&S
®
NGL200/NGM200
38
User Manual 1178.8736.02 ─ 06
[Back] key
Using the [Back] key, you can do several things:
●
Navigate to the previous menu window
●
Close or discard changes made on the on-screen keypad
●
Close the instrument pop-up messages
5.3.3
Output and Channel Controls
Applicable only for 2-channel models (NGL202, NGM202), these keys control the
channel output settings of the instrument.
Function keys
Description
[Ch1], [Ch2]
Selects the respective channel for output.
[Output]
Master output switch - it turns output for all
selected channels on or off.
5.4
Power Derating
The NGL202, NGM202 models include two identical channels with a continuous volt-
age range of 0 V to 20.05 V. The instrument provides a source of up to 6.01 A for volt-
age below 6 V and 3.01 A for voltage range from 6 V to 20 V.
Similar to NGL202, NGM202, the NGL201, NGM201 models provide a single channel
with an output power of up to 60 W.
Combination of the set voltage and current limit results in the following output perfor-
mance graph.
Figure 5-11: Output performance graph
5.5
Operation Modes
The R&S
NGL/NGM operates in three modes, i.e. the constant voltage (CV), constant
current (CC) and constant resistance (CR). The instrument switches automatically
Operation Modes