Remote Control
R&S
®
NGE100B
90
User Manual 5601.1343.02 ─ 02
Parameters:
<Voltage>
{<Voltage> | MINimum | MAXimum | DEFault}
<Voltage>
0.0 V to 32.2 V (adjustable in 10 mV steps)
MINimum
0.000E+00
MAXimum
32.200E+01
Default unit: V
Example:
VOLT 10
Set the voltage value to 10 V.
VOLT? -> 10
Return the voltage value of the selected channel.
6.5.12
EasyRamp Commands
[SOURce:]VOLTage:EASYramp[:STATe]
<arg0>
This command activates/deactivates or queries the EasyRamp state for the previous
selected channel.
See also
SOURce:VOLTage:RAMP:STATe
.
Parameters:
<arg0>
[SOURce:]VOLTage:EASYramp:TIME
<Time>
This command sets/queries the EasyRamp time of the previous selected channel in
milliseconds. 10 ms to 10 s in steps of 10 ms.
See also
SOURce:VOLTage:RAMP:DURation
.
Parameters:
<Time>
Range:
10 to 10000
*RST:
10
Default unit: ms
[SOURce:]VOLTage:RAMP:DURation
<Time>
This command sets/queries the EasyRamp duration of the previous selected channel
in mili seconds. 10ms to 10 seconds in steps of 10 ms.
Parameters:
<Time>
Range:
10 to 10000
*RST:
10
Default unit: ms
Remote Control - Commands