![R&S NGE-K101 User Manual Download Page 83](http://html1.mh-extra.com/html/rands/nge-k101/nge-k101_user-manual_3088109083.webp)
Remote Control
R&S
®
NGE100B
81
User Manual 5601.1343.02 ─ 02
Example:
INST OUT1
OUTP ON
Ch 1 and output state will be activated; i.e.
Ch 1
and
Output
key
on the front panel will be illuminated.
OUTP? -> 1
Return the output state.
OUTPut:GENeral
<arg0>
This command activates or queries the instrument master output state.
If the master output state is activated, the
Output
key on the front panel will be illumi-
nated.
Parameters:
<arg0>
{OFF | ON | 0 | 1}
ON | 1
Output state will be activated
OFF | 0
Output state will be deactivated
Example:
INST OUT1
OUTP:SEL ON
Volt 12
Curr 0.1
Ch 1 is activated.
INST OUT2
OUTP:SEL ON
Volt 12
Curr 0.2
Ch 2 is activated.
OUTP:GEN ON
Ch 1
and
Ch 2
and
Output
state are activated.
OUTP:GEN? -> 1
Return output state of the channel.
OUTPut:MODE?
Queries the output operating mode of the previous selected channel
Return values:
<arg0>
{OFF | CV | CC}
OFF
Output is turned off.
CV
Output mode changed to constant voltage mode.
CC
Output mode changed to constant current mode.
*RST:
OFF
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH