Quick Start
R&S
®
NGE100B
16
User Manual 5601.1343.02 ─ 02
Use only the power cord included in the delivery package. See
on page 17.
Before each measurement, measuring cables must be inspected for damage and
replaced if necessary. Damaged or worn components can damage the instrument or
cause injury.
The product may be operated only under the operating conditions and in the positions
specified by the manufacturer, without the product's ventilation being obstructed. If the
manufacturer’s specifications are not observed, this can result in electric shock, fire
and/or serious personal injury, and in some cases, death.
Applicable local or national safety regulations and rules for the prevention of accidents
must be observed in all work performed.
The instrument is designed for use in the following sectors: Industrial, residential, busi-
ness and commercial areas and small businesses.
The instrument is designed for indoor use only. Before each measurement, you need
to verify at a known source if the instrument functions properly.
To disconnect from the mains, the low-heat device socket on the back panel has to be
unplugged.
See
for the general data on the instrument specification. For more informa-
tion, see the instrument product brochure (PN: 5214.8748.12).
Table 4-1: General data on instrument specification
Mains nominal voltage
AC
115/230 V (
±
10 %) 50/60 Hz
Power consumption
Maximum input power
180 W
Mains fuses
115 V AC
IEC 60127-2/5 T 5 A 250 V
230 V AC
IEC 60127-2/5 T 2.5 A 250 V
Temperature
Operating temperature range
0
°
C to + 40
°
C
Storage temperature range
- 20
°
C to + 70
°
C
Humidity
Non-condensing
5 % to 80 %
Display
3.5
"
(QVGA)
Rack mount capability
½
19
"
R&S HZC95 option
Dimensions
W x H x D
222 mm x 88 mm x 280 mm
(8.74 in x 3.46 in x 11.02 in)
Weight
R&S NGE102B
4.9 kg (10.80 lb)
R&S NGE103B
5.0 kg (11.02 lb)
Putting into Operation