Instrument Functions
R&S
®
NGE100B
47
User Manual 5601.1343.02 ─ 02
Trigger out parameters Trigger conditions
Description
Output Off
Output the selected logic level when the
Output
is turned off at the selected channel.
Fuse Tripped
Output the selected logic level when a fuse
tripped event occurs on the selected chan-
nel.
Chapter 5.4.3, "Activating Fuse"
CC Mode
Output the selected logic level when the
selected channel operates in the CC mode.
See
Voltage Level
>= "set value"
Output the selected logic level when the
voltage level of the selected channel is
greater or equal to the set level.
Current Level
>= "set value"
Output the selected logic level when the
current level of the selected channel is
greater or equal to the set level.
Critical Event
OVP
Output the selected logic level when the
selected critical event (OVP, OPP, OTP)
occurs on the selected channel. See
OPP
OTP
EasyArb Active
Output the selected logic level when the
EasyArb function is running on channel 1.
See
Note: For a reliable DIO trigger output, the
EasyArb waveform needs to be at least 200
ms.
Logic
Active High
Set the logic level of the trigger out signal.
Active Low
1. Select "Interface > Digital IO" menu item.
The R&S NGE100B displays the "Digital IO" dialog. See
Figure 5-19: Digital IO dialog
2. Select the desired DIO interface.
Advanced Functions