Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
72
Operating Manual 1176.7610.02 ─ 02
Example:
For the first steps refer to
chapter 4.14, "Programming Exam-
CONF:BTO:MEAS CFST
Activates the Carrier Frequency Stability and Modulation Accu-
racy measurement.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:CFST:DEVM? AVER
Reads the root mean square of the average differential error
vector magnitude.
Usage:
Query only
Mode:
BT
Manual operation:
See
CALCulate<n>:BTOoth:CFSTability:DEVM:DPCT
CALCulate<n>:BTOoth:CFSTability:DEVM:D99Pct?
This command reads 99% of the differential error vector magnitude (DEVM) in the Car-
rier Frequency Stability and Modulation Accuracy measurement.
Note:
This command is only available with active Carrier Frequency Stability and Modulation
Accuracy measurement (see
With all other measurements this command causes a query error.
Suffix:
<n>
.
irrelevant
Example:
For the first steps refer to
chapter 4.14, "Programming Exam-
CONF:BTO:MEAS CFST
Activates the Carrier Frequency Stability and Modulation Accu-
racy measurement.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:CFST:DEVM:D99Pct?
Reads 99% of the differential error vector magnitude.
Usage:
Query only
Mode:
BT
Manual operation:
See
CALCulate<n>:BTOoth:CFSTability:DEVM:PEAK?
This command reads the peak of the differential error vector magnitude (DEVM) in the
Carrier Frequency Stability and Modulation Accuracy measurement.
Note:
CALCulate subsystem