Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
113
Operating Manual 1176.7610.02 ─ 02
This command has the same effect as the
[SENSe<source>:]SWEep:COUNt
com-
mand. In both cases, the number of measurements is defined whether the average cal-
culation is active or not.
The number of measurements applies to all traces in the window.
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<NoMeasurements> 0 to 32767
*RST:
0
Example:
SWE:CONT OFF
Switching to single sweep mode.
AVER:COUN 16
Sets the number of measurements to 16.
AVER:STAT ON
Switches on the calculation of average.
INIT;*WAI
Starts the measurement and waits for the end of the 16 sweeps.
[SENSe:]AVERage<n>:TYPE
<FunctionType>
This command selects the type of average function.
Suffix:
<n>
.
Selects the measurement window.
Parameters:
<FunctionType>
VIDeo | LINear | POWer
VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST:
VIDeo
Example:
AVER:TYPE LIN
Switches to linear average calculation.
[SENSe:]BANDwidth|BWIDth[:RESolution]
<Bandwidth>
This command defines the resolution bandwidth.
The available resolution bandwidths are specified in the data sheet. For details on the
correlation between resolution bandwidth and filter type refer to
"Selecting the Appropriate Filter Type"
SENSe Subsystem (BLUETOOTH, K8)