Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
92
Operating Manual 1176.7610.02 ─ 02
Parameters:
<NoChannels>
1 to 78
*RST:
78
Example:
CONF:BTO:ACLR:ACP 10
Selects 10 adjacent channel pairs.
Mode:
BT
Manual operation:
See
"No. of ACP Chan (TX Spec ACLR, Spurious Emissions)"
CONFigure:BTOoth:BANDwidth|BWIDth[:RESolution]
<Bandwidth>
This command sets the resolution bandwidth of the analyzer for the currently selected
Bluetooth measurement (see
For the measurements of modulation characteristics, initial carrier frequency tolerance,
and carrier frequency drift (see
the resolution bandwidth corresponds to the IF bandwidth of the signal. If the resolution
bandwidth is changed, its coupling to the settings according to the RF Test Specifica-
tion is cancelled.
Note:
The settings for the measurements are valid for all three measurements.For all other
Bluetooth measurements the setting is only valid for the currently active measurement
(see
on page 96).This command is not availa-
ble with active Adjacent Channel Power measurement.
Parameters:
<Bandwidth>
300kHz to 10MHz
*RST:
AUTO is set to ON
Example:
INST:SEL BTO
Activates the Bluetooth Measurements option.
CONF:BTO:MEAS OPOW
Activates the Output Power measurement.
CONF:BTO:BAND 1KHZ
Sets the resolution bandwidth to 1kHz.
Mode:
BT
Manual operation:
See
"Res BW Manual (Output Power, Rel TX Power)"
CONFigure:BTOoth:BANDwidth|BWIDth[:RESolution]:AUTO
<Mode>
This command couples the resolution bandwidth to the settings according to the RF
Test Specification.
Note:
CONFigure:BTOoth Subsystem (BLUETOOTH, K8)