Instrument Functions for Bluetooth Measurements (K8)
R&S
®
FSV-K8
54
Operating Manual 1176.7610.02 ─ 02
Video BW Manual (TX Spec ACLR)
..............................................................................54
Video BW Auto (TX Spec ACLR, Spurious Emissions)
................................................ 55
Filter Type (Output Power, TX Spec ACLR, Spurious Emissions, Rel TX Power)
Meas Filter (Modulation Char, Init Carr Freq Tol, Carr Freq Drift)
................................ 55
Sweeptime Manual (TX Spec ACLR)
Opens an edit dialog box to enter the sweep time. The valid value range is 10
μ
s (mini-
mum measurement time for one channel) to 16000 s. The default setting is 79 s.
Remote command:
Sweeptime Auto (TX Spec ACLR)
Activates the automatic calculation of the sweep time. The automatic sweep time cor-
responds to the settings defined in the RF Test Specification. By default, it is activated.
Remote command:
Meas Time Manual
Opens an edit dialog box to enter the measurement time. The valid value range is 1
μ
s
to (400001 / points per symbol)
μ
s.
This function is not available for Spurious Emissions measurements.
Remote command:
Meas Time Auto
Activates the automatic calculation of the measurement time. The automatic sweep
time corresponds to the settings defined in the RF Test Specification. By default, the
automatic sweep time calculation is activated.
This function is not available for Spurious Emissions measurements.
Remote command:
Res BW Manual (Output Power, Rel TX Power)
Opens an edit dialog box to enter the resolution bandwidth. The values range from 1
MHz to 3 MHz. The default value is 3 MHz.
Remote command:
Res BW Auto (Output Power, TX Spec ACLR, Rel TX Power, Spurious Emissions)
Sets the bandwidth according to the values defined in the RF Test Specification.
Remote command:
Video BW Manual (TX Spec ACLR)
The default value is 300 kHz, according to the values defined in the RF Test Specifica-
tion.
Softkeys of the Bandwidth Menu (option R&S
FSV-K8)