Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
83
Operating Manual 1176.7610.02 ─ 02
Mode:
BT
Manual operation:
See
CALCulate<n>:BTOoth:OPOWer[:PEAK]?
This command reads the peak value of the Output Power measurement according to
the BLUETOOTH standard.
Note:
This command is only available with active Output Power measurement (see
on page 96). With all other measurements it
causes a query error.
Suffix:
<n>
.
irrelevant
Example:
For the first steps refer to
chapter 4.14, "Programming Exam-
CONF:BTO:MEAS OPOW
Activates the Output Power measurement.
INIT;*WAI
Starts the measurement with synchronization.
CALC:BTO:OPOW?
Queries the output power result.
Usage:
Query only
Mode:
BT
Manual operation:
See
CALCulate<n>:BTOoth:OPOWer:AVERage?
<Mode>
This command reads the average value of the Output Power measurement according
to the Bluetooth standard.
With a sweep count value of
≥
) and trace
mode "Clear/Write" (
DISPlay[:WINDow<n>]:TRACe<t>:MODE
) the selected num-
ber of measurements is performed when a single sweep is started (
). During these measurements the minimum and maximum values are
determined. If only a single measurement is performed, the minimum and maximum
value are identical.
Note:
This command is only available with active Output Power measurement (see
on page 96). With all other measurements it
causes a query error.
Suffix:
<n>
.
irrelevant
CALCulate subsystem