Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
110
Operating Manual 1176.7610.02 ─ 02
Example:
INIT:CONT OFF
Switches the sequence to single sweep.
INIT:CONT ON
Switches the sequence to continuous sweep.
Mode:
all
Manual operation:
See
See
INITiate<n>[:IMMediate]
The command initiates a new measurement sequence.
With sweep count > 0 or average count > 0, this means a restart of the indicated num-
ber of measurements. With trace functions MAXHold, MINHold and AVERage, the pre-
vious results are reset on restarting the measurement.
In single sweep mode, you can synchronize to the end of the measurement with *OPC,
*OPC? or *WAI. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Suffix:
<n>
.
irrelevant
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
Mode:
all
4.6 INPut subsystem
...............................................................................................111
INPut:ATTenuation
<Value>
This command programs the input attenuator. To protect the input mixer against dam-
age from overloads, the setting 0 dB can be obtained by entering numerals, not by
using the DOWN command.
INPut subsystem