Remote Commands for Bluetooth Measurements (K8)
R&S
®
FSV-K8
112
Operating Manual 1176.7610.02 ─ 02
INPut:IMPedance
<Impedance>
This command selects the nominal input impedance.
75
Ω
should be selected if the 50
Ω
input impedance is transformed to a higher impe-
dance using a 75
Ω
adapter of the RAZ type (= 25
Ω
in series to the input impedance
of the instrument). The correction value in this case is 1.76 dB = 10 log (75
Ω
/50
Ω
).
This function is not available if the R&S Digital I/Q Interface (R&S
FSV-B17) is active.
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Example:
INP:IMP 75
Manual operation:
See
4.7 INSTrument Subsystem (BLUETOOTH, K8)
INSTrument[:SELect]
Parameters:
Parameter
BTOoth
Bluetooth mode (option Bluetooth Measurements)
INSTrument:NSELect
Parameters:
Parameter
12
Bluetooth mode (option Bluetooth Measurements)
4.8 SENSe Subsystem (BLUETOOTH, K8)
[SENSe:]AVERage<n>:COUNt
<NoMeasurements>
This command defines the number of measurements which contribute to the average
value.
Note that continuous averaging is performed after the indicated number has been
reached in continuous sweep mode.
In single sweep mode, the sweep is stopped as soon as the indicated number of meas-
urements (sweeps) is reached. Synchronization to the end of the indicated number of
measurements is only possible in single sweep mode.
INSTrument Subsystem (BLUETOOTH, K8)