![R&S 1444.1190K02 User Manual Download Page 40](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881040.webp)
Remote control commands
R&S
®
NRPxxP
40
User Manual 1179.5760.02 ─ 02
8.4.2
Configuring the measured values
Before starting a measurement, you can configure the measurand or enable the mea-
surement of additional-measured values.
............................................................................................... 41
CALCulate:FEED
<mode>
If you query measurement data using
FETCh<Sensor>[:SCALar][:POWer][:
, the power sensor returns data of the measurand that was configured before.
Generally, this measurand is the average power. However, the power sensor can also
output data of other measurands.
To configure which measurand the
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
command reads, use the
CALCulate:FEED
command before the measurement is ini-
tiated. Depending on the measurement mode, the following settings are possible:
SENS:FUNC
Possible CALC:FEED
Meaning
"POWer:AVG"
"POWer:AVERage"
Average value
"POWer:PEAK"
Peak value
"POWer:RANDom"
Randomly selected value from the
measurement interval
"POWer:BURSt:AVG"
"POWer:AVERage"
Average value
"POWer:PEAK"
Peak value
"POWer:RANDom"
Randomly selected value from the
measurement interval
"POWer:TSLot:AVG"
"POWer:AVERage"
Average value
"POWer:PEAK"
Peak value
"POWer:RANDom"
Randomly selected value from the
measurement interval
"XTIMe:POWer"
"POWer:TRACe"
Measurement sequence
"POWer:PEAK:TRACe"
Peak value of the samples per
trace point
"POWer:RANDom:TRACe"
Randomly selected value of the
samples per trace point
"XPOWer:CCDFunction"
CCDF:TRACe
Complementary cumulative distri-
bution function
"XPOWer:PDFunction"
PDF:TRACe
Probability density function
Parameters:
<mode>
*RST:
"POWer:AVERage"
Preparing for the measurement