Remote control commands
R&S
®
NRPxxP
116
User Manual 1179.5760.02 ─ 02
SYSTem:RESTart
Restarts the firmware of the power sensor.
Usage:
Event
SYSTem:RUTime
<update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode,
Sets the result update time. That is the maximum rate in which the power sensor can
output measurement results.
Parameters:
<update_time>
Range:
0.0 to 10.0
*RST:
0.1
Default unit: Seconds
SYSTem:SERRor?
Queries the next static error, if available.
Static errors, as a rule, prevent the execution of normal measurements.
Errors in SCPI communication are queried using
.
Usage:
Query only
SYSTem:SUTime
<update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode,
Sets the status update time. That is the maximum rate in which the power sensor can
output measurement results.
Parameters:
<update_time>
Range:
0.0 to 10.0
*RST:
10e-3
Default unit: Seconds
SYSTem:TLEVels?
Queries the possible power test levels of the power sensor.
Usage:
Query only
SYSTem:TRANsaction:BEGin
Starts a series of settings.
Configuring the system