![R&S 1444.1190K02 User Manual Download Page 18](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881018.webp)
Power sensor tour
R&S
®
NRPxxP
18
User Manual 1179.5760.02 ─ 02
4
Power sensor tour
This chapter provides an overview of the available connectors and LEDs of the power
sensor.
NRP
18P
SM
AR
T SENSOR
TECHNOLOGY
1
4
2
3
Figure 4-1: R&S
NRPxxP power sensors (example)
1 = RF connector, see
2 = Status LED, see
Chapter 4.2, "Status information"
3 = Host interface connector, see
4 = Trigger I/O connector, see
Chapter 4.4, "Trigger I/O connector"
4.1
RF connector
The RF connector is used for connecting the power sensor to a device under test
(DUT) or a signal generator. See
Chapter 3.4, "Connecting to a DUT"
For maximum measurement accuracy, tighten the RF connector using a torque wrench
with a nominal torque as specified in the following table.
Table 4-1: R&S
NRPxxP RF connector characteristics
Power sensor
Male connector
Matching female con-
nector
Tightening torque
R&S
NRP18P
N
N
1.36
Nm (12''
lbs)
R&S
NRP40P
2.92
mm
3.50
mm/ 2.92
mm/ SMA
0.90
Nm (8''
lbs)
R&S
NRP50P
2.4
mm
2.4
mm/ 1.85
mm
RF connector