Remote control commands
R&S
®
NRPxxP
99
User Manual 1179.5760.02 ─ 02
The fast Fourier transform (FFT) of the measured trace data is based on the following
parameters:
●
Window function
Set by
SENSe<Sensor>:TRACe:FFT:WINDow
.
●
Number of values used
The number of required values per trace sequence is set by
. For computing the FFT, the closest power of two below this num-
ber is used, 2
N
. For example, if the set number is 300, then 256 values are used in
the FFT, 2
8
.
●
Number of values returned
The number of computed values is half the number of used values.
●
Distance of the spectral lines in Hz
Equals the reciprocal of the trace length, which is set by
Note:
The spectrum computed as described is the spectrum of the
envelope power
,
different from the results displayed by a spectrum analyzer. For questions on details,
contact our support team.
Usage:
Query only
8.9.5
Pulse analysis measurement results
For information on the pulse analysis and how to configure the measurement, see
Chapter 8.6.5, "Pulse analysis measurement"
Querying measurement results