![R&S 1444.1190K02 User Manual Download Page 64](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881064.webp)
Remote control commands
R&S
®
NRPxxP
64
User Manual 1179.5760.02 ─ 02
MOVing
Outputs intermediate values to facilitate early detection of
changes in the measured quantity. In the settled state, that
means when the number of measurements specified by the
average count has been performed, a moving average is output.
REPeat
Specifies that a measurement result is not output until the entire
measurement has been completed. This means that the number
of measurement cycle repetitions is equal to the set average
count. If the average count is large, the measurement time can
be very long.
The average count is set using
*RST:
REPeat
Example:
TRAC:AVER:TCON REP
SENSe<Sensor>:TRACe:AVERage[:STATe]
<state>
Enables or disables the averaging filter.
Parameters:
<state>
*RST:
ON
SENSe<Sensor>:TRACe:FFT:WINDow
<window>
Selects a window function for computing a fast Fourier transform (FFT) of the trace.
The resulting power spectrum of the trace is queried by
.
1.0
0.0
-1.0
1.0
4
3
2
1
samples
amplitude
1 = rectangular window
2 = Hamming window
3 = Hanning window
4 = Blackman window
Configuring the measurement modes